Pain Panel from Urine using Strata DE SLE and a Kinetex Biphenyl Core-Shell LC Column

Pain Panel from Urine using Strata DE SLE and a Kinetex Biphenyl Core-Shell LC Column
LC Conditions
Column

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4622-AN

Phase Name

Biphenyl

Elution Type

Gradient

Mobile Phase

A: 0.1% formic acid in water

B: 0.1% formic acid in acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

4

0

100

3

5

0

100

4

5.01

95

5

5

7.1

95

5

Flow Rate

500.00 mL/min

Temperature

22°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4622-AN

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. morphine
  • 2. moprhine-D3
  • 3. hydromorphone
  • 4. codeine D6
  • 5. codeine
  • 6. methamphetamine
  • 7. oxycodone
  • 8. oxycodone D6
  • 9. hydrocodone D6
  • 10. hydrocodone
  • 11. norfentanyl
  • 12. benzoylecognine
  • 13. norbuprenorphine
  • 14. PCP
  • 15. fentanyl
  • 16. buprenorphine
  • 17. nordiazepam
  • 18. lorazepam
  • 19. lorazepam D4
  • 20. alprazolam
  • 21. diazepam
Similar Applications