IntantPC Labeled N-Linked Glycans from Infliximab-dyyb on bioZen Glycan
LC Conditions
Column
bioZen 2.6 µm Glycan, LC Column 150 x 2.1 mm, Ea
Brand
Biozen LC
Part No
00F-4773-AN
Phase Name
Glycan
Elution Type
Gradient
Mobile Phase
A: 100 mM Ammonium Formate, pH 4.5
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
22
78
2
10
25.5
74.5
3
24
28
72
4
38.5
44.1
55.9
5
38.6
60
40
6
40.6
60
40
7
40.7
22
78
8
48
22
78
Flow Rate
0.50 mL/min
Temperature
50°C
System
Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC H-Class
Detection
LC/FLD (Fluorescence Detector) @ 285 nm (nanometers)
Detection Info
Sample Notes
Utilize bioZen N-Glycan Clean-Up microelution SPE 96well plate for sample preparation (Part No.: 8M-S009-NGA).
Order Items Used in This Application
- 1. G0-GlcNac
- 2. G1-GlcNac
- 3. G0
- 4. Man5
- 5. G0F
- 6. G1F-GlcNac
- 7. G1F
- 8. G1F'
- 9. G1FB
- 10. G2F
- 11. G2FB
- 12. Sialylated Glycan
- 13. Sialylated Glycan
- 14. Sialylated Glycan
- 15. Sialylated Glycan
- 16. Sialylated Glycan
- 17. Sialylated Glycan
- 18. Sialylated Glycan
- 19. Sialylated Glycan
- 20. Sialylated Glycan
- 21. Sialylated Glycan
- 22. Sialylated Glycan
Similar Applications