Charge Variants of Cetuximab Biosimilar using pH Gradient on bioZen 6 µm WCX

Charge Variants of Cetuximab Biosimilar using pH Gradient on bioZen 6 µm WCX
LC Conditions
Column

bioZen 6µm WCX, LC Column 250 x 4.6 mm, Ea

Brand

Biozen LC

Part No

00G-4777-E0

Phase Name

WCX

Elution Type

Gradient

Mobile Phase

A: CX-1 , pH 5.6

B: CX-1, pH 10.2

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

100

0

2

1

100

0

3

21

0

100

4

23

0

100

7

24

100

0

Flow Rate

1.00 mL/min

Temperature

30°C

System

Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC H-Class

Detection

LC/UV (DAD, PDA) @ 280 nm (nanometers)

Detection Info

Order Items Used in This Application

bioZen 6µm WCX, LC Column 250 x 4.6 mm, Ea
00G-4777-E0

bioZen 6µm WCX, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Cetuximab
Similar Applications