Nimesulide in formulation gel using Kinetex 2.6µm Polar C18 150x.4.6mm
LC Conditions
Column
Kinetex 2.6 µm Polar C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Kinetex
Part No
00F-4759-E0
Phase Name
Polar C18
Guard Cartridge:
AJ0-9530
Guard Holder:
AJ0-9000
Elution Type
Gradient
Mobile Phase
A: 30mM Potassium Phosphate pH 3.0
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
50
50
2
8
20
80
3
9
20
80
4
10
50
50
5
12
50
50
Flow Rate
1.00 mL/min
Temperature
40°C
System
Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC I-Class
Detection
LC/UV (DAD, PDA) @ 254 nm (nanometers) (30°C)
Detection Info
Sample Notes
Gel formulation extraction
Order Items Used in This Application
- 1. Excepient
- 2. Nimesulide
Similar Applications
No Similar applications found.