Albuterol and Ambroxol in Oral Solution using Kinetex 2.6µm Polar C18 150x.4.6mm

Albuterol and Ambroxol in Oral Solution using Kinetex 2.6µm Polar C18 150x.4.6mm
LC Conditions
Column

Kinetex 2.6 µm Polar C18 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Kinetex

Part No

00F-4759-E0

Phase Name

Polar C18

Guard Cartridge:

AJ0-9530

Guard Holder:

AJ0-9000

Elution Type

Gradient

Mobile Phase

A: 30mM Potassium Phosphate pH 3.0

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

100

0

2

1

100

0

3

7

40

60

4

9

40

60

5

10

100

0

6

13

100

0

Flow Rate

1.00 mL/min

Temperature

40°C

System

Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC I-Class

Detection

LC/UV (DAD, PDA) @ 254 nm (nanometers) (30°C)

Detection Info
Sample Notes

Gel formulation extraction

Order Items Used in This Application

SecurityGuard ULTRA Holder, for UHPLC Columns 2.1 to 4.6mm ID, Ea
AJ0-9000

SecurityGuard ULTRA Holder, for UHPLC Columns 2.1 to 4.6mm ID, Ea

View Product

SecurityGuard ULTRA Cartridges, UHPLC Polar C18 4.6mm ID Columns, 3/Pk
AJ0-9530

SecurityGuard ULTRA Cartridges, UHPLC Polar C18 4.6mm ID Columns, 3/Pk

View Product

Kinetex 2.6 µm Polar C18 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4759-E0

Kinetex 2.6 µm Polar C18 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Albuterol
  • 2. Ambroxol
Similar Applications

No Similar applications found.