Benzalkonium Chloride UPS method using Kinetex 2.6u Biphenyl 100x2.1mm
LC Conditions
Column
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 100 x 2.1 mm, Ea
Brand
Kinetex
Part No
00D-4622-AN
Phase Name
Biphenyl
Guard Cartridge:
AJ0-4304
Guard Holder:
KJ0-4282
Elution Type
Isocratic
Mobile Phase
A: Water with 100 mM Ammonium Acetate and 0.2 % Phosphoric Acid
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
5
35
65
Flow Rate
0.30 mL/min
Temperature
30°C
System
Waters ACQUITY UPLC I-Class Waters ACQUITY UPLC I-Class
Detection
LC/UV (DAD, PDA) @ 214 nm (nanometers) (20°C)
Detection Info
Syringe Filter
AF0-2207-12
Order Items Used in This Application
- 1. C12
- 2. C14
- 3. C16
Similar Applications
No Similar applications found.