16 Cannabinoids on Kinetex C18
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Kinetex
Part No
00F-4462-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% formic acid in water
B: 0.1% formic acid in 75:25 Methanol:Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
20
80
2
8
20
80
3
10
0
100
4
12
0
100
Flow Rate
1.50 mL/min
Temperature
30°C
System
1200 Series high performance autosampler, 0.1 100 µL Agilent Technologies 1200 Series
Detection
LC/UV (DAD, PDA) @ 228
Detection Info
Order Items Used in This Application
- 1. CBDV
- 2. CBDVA
- 3. CBD
- 4. CBG
- 5. THCV
- 6. CBDA
- 7. CBGA
- 8. CBN
- 9. D9-THC
- 10. THCVA
- 11. D8-THC
- 12. CBL
- 13. CBC
- 14. CBNA
- 15. THCA-A
- 16. CBCA
Similar Applications