Paraffin Marker Standards on ZB-DHA-PONA 100m x 0.25mm x 0.5um
GC Conditions
Column
Zebron ZB-DHA-PONA, GC Cap. Column 100m x 0.25mm x 0.5µm, Ea
Part No
7MG-G042-17
Phase Name
ZB-DHA-PONA
Injection
Split 40:1 @ 300°C
Inlet Liner
AG2-0A11-05
Oven Program
35°C for 14 min to 60°C @ 1.1°C/min 60°C for 19 min to 280°C @ 2°C/min 280°C for 5 min
Carrier Gas
Constant Flow Helium, 1.6 mL/min
System
Agilent Technologies 6890 HP 6890
Detection
GC/FID @ (320°C)
Detection Info
Sample Notes
The pariffins standard was first diluted 100:1 in pentane and then run with a 40:1 split to avoid column overloading.
Order Items Used in This Application
AG2-0A11-05
Zebron PLUS Liner for Agilent & Thermo, 4mm ID Single Taper Wool on Bottom, 5/pk
- 1. Pentane
- 2. Hexane
- 3. Heptane
- 4. Octane
- 5. Nonane
- 6. Decane
- 7. Undecane
- 8. Dodecane
- 9. Tridecane
- 10. Tetradecane
- 11. Pentadecane