Olefins Marker Standards on ZB-DHA-PONA 100m x 0.25mm x 0.5um
GC Conditions
Column
Zebron ZB-DHA-PONA, GC Cap. Column 100m x 0.25mm x 0.5µm, Ea
Part No
7MG-G042-17
Phase Name
ZB-DHA-PONA
Injection
Split 40:1 @ 300°C
Inlet Liner
AG2-0A11-05
Oven Program
35°C for 14 min to 60°C @ 1.1°C/min 60°C for 19 min to 280°C @ 2°C/min 280°C for 5 min
Carrier Gas
Constant Flow Helium, 1.6 mL/min
System
Agilent Technologies 6890 HP 6890
Detection
GC/FID @ (320°C)
Detection Info
Sample Notes
The pariffins standard was first diluted 100:1 in pentane and then run with a 40:1 split to avoid column overloading.
Order Items Used in This Application
AG2-0A11-05
Zebron PLUS Liner for Agilent & Thermo, 4mm ID Single Taper Wool on Bottom, 5/pk
- 1. 3-Methyl-1-butene
- 2. 1-Pentene
- 3. 2-Methyl-1-butene
- 4. 2-Methyl-1,3-butadiene
- 5. trans-2-Pentene
- 6. cis-2-Pentene
- 7. 4-Methylpentene-1
- 8. 1-Hexene
- 9. 2-Methylpentene-2
- 10. trans-2-Hexene
- 11. cis-2-Hexene
- 12. 1-Heptene
- 13. trans-3-Heptene
- 14. cis-3-Heptene
- 15. trans-2-Heptene
- 16. cis-2-Heptene
- 17. 1-Octene
- 18. trans-2-Octene
- 19. cis-2-Octene
- 20. 1-Nonene
- 21. trans-3-Nonene
- 22. cis-3-Nonene
- 23. trans-2-Nonene
- 24. cis-2-Nonene
- 25. 1-Decene
Similar Applications