Lux i-Amylose-1 Separation for Thalidomide

Lux i-Amylose-1 Separation for Thalidomide
LC Conditions
Column

Lux 5 µm i-Amylose-1, LC Column 250 x 4.6 mm, Ea

Brand

Lux

Part No

00G-4762-E0

Phase Name

i-Amylose-1

Elution Type

Isocratic

Mobile Phase

A: Acetonitrile

B: Diethylamine

Gradient Profile

Step No.

Time(min)

%A

%B

1

8

100

0.1

Flow Rate

1.00 mL/min

Temperature

25°C

System

Agilent Technologies Agilent Technologies 1100 Series

Detection

LC/UV (DAD, PDA) @ 254 nm (nanometers)

Detection Info
Sample Notes

2 mg/mL in Ethanol

Order Items Used in This Application

Lux 5 µm i-Amylose-1, LC Column 250 x 4.6 mm, Ea
00G-4762-E0

Lux 5 µm i-Amylose-1, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Thalidomide
-OEChem-06211003152D<br><br> 29 31  0     1  0  0  0  0  0999 V2000<br>    6.2619   -1.7320    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.9889    1.7552    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.9889   -1.7552    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    9.2619    0.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.2619    0.0000    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    7.7619   -0.8660    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.2619    0.0000    0.0000 C   0  0  3  0  0  0  0  0  0  0  0  0<br>    6.7619    0.8660    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    7.7619    0.8660    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.7619   -0.8660    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.6783    0.8047    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.6783   -0.8047    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    8.2619    0.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660    1.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -1.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000    0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.9519   -0.5369    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.8695    1.4766    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.1793    1.0781    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    8.3445    1.0781    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    7.6542    1.4766    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    8.0719   -1.4030    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660    1.6200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -1.6200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4631    0.8100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4631   -0.8100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1 10  2  0  0  0  0<br>  2 11  2  0  0  0  0<br>  3 12  2  0  0  0  0<br>  4 15  2  0  0  0  0<br>  5  7  1  0  0  0  0<br>  5 11  1  0  0  0  0<br>  5 12  1  0  0  0  0<br>  6 10  1  0  0  0  0<br>  6 15  1  0  0  0  0<br>  6 25  1  0  0  0  0<br>  7  8  1  0  0  0  0<br>  7 10  1  0  0  0  0<br>  7 20  1  0  0  0  0<br>  8  9  1  0  0  0  0<br>  8 21  1  0  0  0  0<br>  8 22  1  0  0  0  0<br>  9 15  1  0  0  0  0<br>  9 23  1  0  0  0  0<br>  9 24  1  0  0  0  0<br> 11 13  1  0  0  0  0<br> 12 14  1  0  0  0  0<br> 13 14  1  0  0  0  0<br> 13 16  2  0  0  0  0<br> 14 17  2  0  0  0  0<br> 16 18  1  0  0  0  0<br> 16 26  1  0  0  0  0<br> 17 19  1  0  0  0  0<br> 17 27  1  0  0  0  0<br> 18 19  2  0  0  0  0<br> 18 28  1  0  0  0  0<br> 19 29  1  0  0  0  0<br>M  END
Compound Name

Thalidomide

CID

5426

Molecular Formula

C13H10N2O4

Molecular Weight

258

No. Hydrogen Bond Acceptors

4

No. Hydrogen Bond Donors

1

Similar Applications

Chiral

Lux i-Cellulose-5 Separation for Thalidomide

Lux 5 µm i-Cellulose-5, LC Column 250 x 4.6 mm, EA

25372

Chiral

Lux i-Amylose-1 Separation for Thalidomide

Lux 5 µm i-Amylose-1, LC Column 250 x 4.6 mm, Ea

25371

Chiral

Lux i-Amylose-3 Separation for Thalidomide

Lux 5 µm i-Amylose-3, LC Column 250 x 4.6 mm, Ea

25342

Thalidomide on Lux 5µm Amylose-2 in NP

20507

Chiral

Thalidomide on Lux 5µm Cellulose-3 in RP

Lux 5 µm Cellulose-3, LC Column 250 x 4.6 mm, Ea

20291

Chiral

Thalidomide on Lux 5µm Cellulose-2 in RP

Lux 5 µm Cellulose-2, LC Column 250 x 4.6 mm, Ea

20239

Chiral

Thalidomide on Lux 5µm Cellulose-4 in RP

Lux 5 µm Cellulose-4, LC Column 250 x 4.6 mm, Ea

20149