GenX replacement PFOA compound by LC-MS/MS
LC Conditions
Column
Gemini 3 µm C18 110 Å, LC Column 50 x 2 mm, Ea
Brand
Gemini
Part No
00B-4439-B0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 20mM Ammonium Acetate in Water
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
0.1
45
55
3
4.5
1
99
4
8
1
99
5
8.5
95
5
Flow Rate
0.60 mL/min
Temperature
40°C
System
Shimadzu SCIEX 4500
Detection
LC/MS/MS
Detection Info
Sample Preparation Details
Description
Strata GCB, 500mg/6mL, 30/Pk
Part No
8B-S528-HCH
Sample Pretreatment
All standards including HFPO-DA and its stable isotope labelled surrogate 13C3-HFPO-DA were purchased from Wellington Laboratories (Guelph, Ontario). 200 mL water samples were extracted using Strata®-XL-AW weak anion exchange SPE cartridges following the conditions in the ISO standard 25101. Samples were also passed through Strata GCB following the requirements of DOD QSM 5.1. The final volume of the SPE eluent was 8 mL. Final extracts consisted of 80% methanol and 20% water. Solid Phase Extraction Protocol Cartridge: Strata XL-AW 500 mg/6 mL Part No.: (8B-S051-HCH) Condition: 6 mL 0.3% Ammonium hydroxide in Methanol, followed by 6 mL Methanol Equilibrate: 6mL Water Load: Add sample at about 10 mL/min Note: use an adapter cap (Part no.: AH0-7191) and large volume sample reservoir (Part No.: AH0-7005) Wash: 2x 6mL Water Elute: 10 mL 0.3% Ammonium hydroxide in Methanol(this 10 mL will elute PFAS/GenX analytes) Load: SPE Eluent through Strata GCB 250 mg/6 mL (8B-S528-HCH) Evvaporate: to dryness and reconstitute to 1.0 mL with Methanol/ Water (96:4)(containing internal standards)
Elution 1
Pass Strata XL-AW Eluent through Strata GCB
Evaporation
Evaporate to dryness
Steps
0
Elute
Pass Strata XL-AW Eluent through Strata GCB
Order Items Used in This Application
- 1. PFBS
- 2. PFBS
- 3. HFPO-DA
- 4. HFPO-DA
- 5. 13C3-HFPO-DA
- 6. PFHxS
- 7. PFHxS
- 8. PFOS
- 9. PFOS