Aloe Emodin, Eugenol Methyl Ether, and Aloins (A and B) from Aloe Standards Screen on Kinetex 2.6u PS C18 50x4.6mm
LC Conditions
Column
Kinetex 2.6 µm PS C18 100 Å, LC Column 50 x 4.6 mm, Ea
Brand
Kinetex
Part No
00B-4780-E0
Phase Name
PS C18
Guard Cartridge:
AJ0-8949
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in MeOH
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
75
25
2
0.5
75
25
3
5
20
80
4
7
20
80
5
7.1
75
25
6
9
75
25
Flow Rate
1.20 mL/min
Temperature
30°C
System
Agilent Technologies 1100 Series Agilent Technologies 1100 Series
Detection
LC/UV (DAD, PDA) @ 220 nm (nanometers) (30°C)
Detection Info
Order Items Used in This Application
- 1. Aloin A
- 2. Aloin B
- 3. Methyl Eugenol
- 4. Aloe Emodin
Similar Applications