Extraction of Vet Drugs in Milk Using Strata-X PRO with a Kinetex 2.6 µm Biphenyl 50 x 3.0mm
LC Conditions
Column
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Brand
Kinetex
Part No
00B-4622-Y0
Phase Name
Biphenyl
Elution Type
Gradient
Mobile Phase
A: 0.1%Formic acid in Water
B: 0.1%Formic acid in Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
1.5
5
95
3
3
5
95
4
3.01
95
5
5
4.5
95
5
Flow Rate
0.50 mL/min
Temperature
45°C
System
Agilent Technologies 1260 Infinity 1200 Series high performance autosampler, 0.1 100 µL
Detection
LC/MS/MS @ 210 nm (nanometers) (30°C)
Detection Info
Sample Notes
ALTA DENA 2% Reduced Fat Milk
Sample Preparation Details
Description
Strata-X PRO, 60mg/3mL, 50/Pk
Part No
8B-S536-UBJ
Sample Pretreatment
To 1 mL of milk add 3 mL of 0.2% formic acid in Acetonitrile /Methanol (90:10). Mix/vortex for 15-20 secs. Centrifuge for 5 min at 10,000 rpm. Collect supernatant to perform SPE in next step.
Load
Supernatant (from the pre-treatment) through cartridge and collect
Evaporation
Evaporate to dryness
Reconstitute
Dried sample in 1 mL of 0.1% Formic acid in Water/ 0.1% Formic acid in Acetonitrile (95:5)
Steps
0
Load
Supernatant (from the pre-treatment) through cartridge and collect
1
Dry Down
Sample under a gentle stream of Nitrogen
2
Reconstitute
Dried sample in 1 mL of 0.1% Formic acid in Water/ 0.1% Formic acid in Acetonitrile (95:5)
Order Items Used in This Application
- 1. Sulfaguanidine
- 2. Lincomycin
- 3. Sulfadiazine
- 4. Cephapirin
- 5. Sulfamerazine
- 6. Sulfamethoxazole
- 7. Sulfamethizole
- 8. Cefalexin
- 9. Sulfamethazine
- 10. Cortisone
- 11. Cortisol
- 12. β-methasone
- 13. Prednisolone
Similar Applications