Calibration Curves for (R)-Amlodipine and (S)-Amlodipine
LC Conditions
Column
Lux 3 µm Cellulose-4, LC Column 150 x 2 mm, Ea
Brand
Lux
Part No
00F-4490-B0
Phase Name
Cellulose-4
Elution Type
Gradient
Mobile Phase
A: 0.05% Ethanolamine in Acetonitrile and Isopropyl Alcohol (96:4 v/v)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
2
4
100
Flow Rate
0.30 mL/min
Temperature
25°C
System
Sciex SCIEX 4500
Detection
LC/MS/MS @
Detection Info
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Pretreatment Note
5 µL of internal standard working solution was added to 100 µL spiked human plasma (spiked at 0.050, 0.150, 20.0, 40.0, and 50.0 ng/mL) and mixed, 500 µL of 0.2% (v/v) Ethanolamine in water was added to sample and vortexed.
Condition 1
1 mL Methanol
Load
Spiked Plasma
Wash 1
1 mL Water
Wash 2
1 mL 20% Methanol in Water
Elution 1
1 mL of 0.1% Formic Acid in Methanol
Evaporation
Evaporate to dryness
Reconstitute
100 µL of Mobile Phase
Steps
0
Condition
1 mL Methanol
1
Equilibrate
1 mL Water
2
Load
Spiked Plasma
3
Wash
1 mL Water
4
Wash
1 mL 20% Methanol in Water
5
Elute
1 mL of 0.1% Formic Acid in Methanol
6
Dry Down
Take to dryness under a gentle stream of Nitrogen at 50 °C
7
Reconstitute
100 µL of Mobile Phase
Order Items Used in This Application