ASTM D5441 (High Concentration Mix-Sigma Aldrich 47942) on ZB-DHA-PONA 50m x 0.20mm x 0.50µm

ASTM D5441 (High Concentration Mix-Sigma Aldrich 47942) on ZB-DHA-PONA 50m x 0.20mm x 0.50µm
GC Conditions
Column

Zebron ZB-DHA-PONA, GC Cap. Column 50m x 0.20mm x 0.5µm, Ea

Part No

7JE-G042-17

Phase Name

ZB-DHA-PONA

Injection

Split 50:1 @ 200°C

Inlet Liner

AG2-0A03-01

Oven Program

40°C for 13 min to 180°C @ 10°C/min 180°C for 3 min

Carrier Gas

Constant Flow Helium, 0.9 mL/min

System

Agilent Technologies 6890 HP 6890/6890

Detection

GC/FID @ (250°C)

Detection Info
Sample Notes

Resolution between trans-2-pentene/tert-butanol and tert-butanol/cis-2-pentene meets the ASTM Method D5441 requirement of R>/=1.3

Order Items Used in This Application

Zebron PLUS Liner for Agilent & Thermo, 4mm ID Straight Z-Liner, EA
AG2-0A03-01

Zebron PLUS Liner for Agilent & Thermo, 4mm ID Straight Z-Liner, EA

View Product

Zebron ZB-DHA-PONA, GC Cap. Column 50m x 0.20mm x 0.5µm, Ea
7JE-G042-17

Zebron ZB-DHA-PONA, GC Cap. Column 50m x 0.20mm x 0.5µm, Ea

View Product

  • 1. Methanol
  • 2. Iso-Pentane
  • 3. n-Pentane
  • 4. Trans-2-Pentene
  • 5. Tert-Butanol
  • 6. Cis-2-Pentene
  • 7. 2-Methyl-2-Butene
  • 8. MTBE
  • 9. Tert-Butyl Ethyl ether
  • 10. Tert-Amyl methyl ether
  • 11. 2,4,4-Trimethyl-1-pentene
  • 12. 4,4�-Dimethyl-2-neopenyl-1-pentene (triisobutylene)
  • 13. 2,2,4,6,6-Pentamethyl-3-heptene (triisobutylene)
Similar Applications