ASTM D5441 (High Concentration Mix-Sigma Aldrich 47942) on ZB-DHA-PONA 50m x 0.20mm x 0.50µm
GC Conditions
Column
Zebron ZB-DHA-PONA, GC Cap. Column 50m x 0.20mm x 0.5µm, Ea
Part No
7JE-G042-17
Phase Name
ZB-DHA-PONA
Injection
Split 50:1 @ 200°C
Inlet Liner
AG2-0A03-01
Oven Program
40°C for 13 min to 180°C @ 10°C/min 180°C for 3 min
Carrier Gas
Constant Flow Helium, 0.9 mL/min
System
Agilent Technologies 6890 HP 6890/6890
Detection
GC/FID @ (250°C)
Detection Info
Sample Notes
Resolution between trans-2-pentene/tert-butanol and tert-butanol/cis-2-pentene meets the ASTM Method D5441 requirement of R>/=1.3
Order Items Used in This Application
- 1. Methanol
- 2. Iso-Pentane
- 3. n-Pentane
- 4. Trans-2-Pentene
- 5. Tert-Butanol
- 6. Cis-2-Pentene
- 7. 2-Methyl-2-Butene
- 8. MTBE
- 9. Tert-Butyl Ethyl ether
- 10. Tert-Amyl methyl ether
- 11. 2,4,4-Trimethyl-1-pentene
- 12. 4,4�-Dimethyl-2-neopenyl-1-pentene (triisobutylene)
- 13. 2,2,4,6,6-Pentamethyl-3-heptene (triisobutylene)