Analysis of EU 15+1 PAHs using a 30 meter Zebron™ ZB-PAH-EU GC Column

Analysis of EU 15+1 PAHs using a 30 meter Zebron™ ZB-PAH-EU GC Column
GC Conditions
Column

Zebron ZB-PAH-EU, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea

Part No

7HG-G043-10

Phase Name

ZB-PAH-EU

Injection

Split 5:1 @ 330°C

Inlet Liner

AG2-4B13-05

Oven Program

45°C for .8 min to 200°C @ 45°C/min to 226°C @ 3°C/min to 320°C @ 10°C/min 320°C for 20 min

Carrier Gas

Constant Pressure Helium, 24.0 psi

System

Shimadzu 2010 Shimadzu GC/MS System

Detection

GC/MS @ (300°C)

Detection Info

Order Items Used in This Application

Zebron PLUS Liner for Shimadzu 2010, 3.4mm ID Single Taper Z-Liner, 5/pk
AG2-4B13-05

Zebron PLUS Liner for Shimadzu 2010, 3.4mm ID Single Taper Z-Liner, 5/pk

View Product

Zebron ZB-PAH-EU, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
7HG-G043-10

Zebron ZB-PAH-EU, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea

View Product

  • 1. Benzo[c]fluorene
  • 2. Benz[a]anthracene
  • 3. Cyclopenta[c,d]pyrene
  • 4. Chrysene
  • 5. 5-Methylchrysene
  • 6. Benzo[b]fluoranthene
  • 7. Benzo[k]fluoranthene
  • 8. Benzo[j]fluoranthene
  • 9. Benzo[a]pyrene
  • 10. Indeno[1,2,3-cd]pyrene
  • 11. Dibenzo[a,h]anthracene
  • 12. Benzo[g,h,i]perylene
  • 13. Dibenzo[a,l]pyrene
  • 14. Dibenzo[a,e]pyrene
  • 15. Dibenzo[a,i]pyrene
  • 16. Dibenzo[a,h]pyrene