Analysis of PAHs in Rubber and Plastic using a 10 meter Zebron™ ZB-EU-PAH GC column
GC Conditions
Column
Zebron ZB-PAH-EU, GC Cap. Column 10 m x 0.10 mm x 0.08 µm, Ea
Part No
7CB-G043-59
Phase Name
ZB-PAH-EU
Injection
Split 5:1 @ 320°C
Inlet Liner
AG2-3B03-05
Oven Program
100°C for 3 min to 200°C @ 60°C/min to 270°C @ 22°C/min to 300°C @ 4.5°C/min to 330°C @ 80°C/min 330°C for .5 min
Carrier Gas
Constant Flow Helium, 0.7 mL/min
System
Shimadzu GC System Shimadzu GC/MS System
Detection
GC/MS @ (330°C)
Detection Info
Order Items Used in This Application
- 1. Naphthalene
- 2. Acenaphthylene
- 3. Acenaphthene
- 4. Fluorene
- 5. Phenanthrene
- 6. Anthracene
- 7. Fluoranthene
- 8. Pyrene
- 9. Benz[a]anthracene
- 10. Chrysene
- 11. Benzo[b]fluoranthene
- 12. Benzo[k]fluoranthene
- 13. Benzo[j]fluoranthene
- 14. Benzo[a]pyrene
- 15. Benzo[e]pyrene
- 16. Indeno[1,2,3-cd]pyrene
- 17. Dibenz[a,h]anthracene
- 18. Benzo[g,h,i]perylene
Similar Applications