Analysis of EU-PAH 15+1 and & Triphenylene PAH using a 30 meter Zebron ZB-PAH-CT GC column

Analysis of EU-PAH 15+1 and & Triphenylene PAH using a 30 meter Zebron ZB-PAH-CT GC column
GC Conditions
Column

Zebron ZB-PAH-CT, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea

Part No

7HG-G044-10

Phase Name

ZB-PAH-CT

Injection

Split 15:1 @ 320°C

Inlet Liner

AG2-4B13-05

Oven Program

45°C for .8 min to 200°C @ 45°C/min to 266°C @ 3°C/min to 320°C @ 10°C/min 320°C for 20 min

Carrier Gas

Constant Pressure Helium, 23.7 psi

System

Shimadzu 2010 Shimadzu GC/MS System

Detection

GC/MS @ (300°C)

Detection Info

Order Items Used in This Application

Zebron PLUS Liner for Shimadzu 2010, 3.4mm ID Single Taper Z-Liner, 5/pk
AG2-4B13-05

Zebron PLUS Liner for Shimadzu 2010, 3.4mm ID Single Taper Z-Liner, 5/pk

View Product

Zebron ZB-PAH-CT, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
7HG-G044-10

Zebron ZB-PAH-CT, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea

View Product

  • 1. Benzo[c]fluorene
  • 2. Cyclopenta[c,d]pyrene
  • 3. Benz[a]anthracene
  • 4. Triphenylene
  • 5. Chrysene
  • 6. 5-Methylchrysene
  • 7. Benzo[b]fluoranthene
  • 8. Benzo[j]fluoranthene
  • 9. Benzo[k]fluoranthene
  • 10. Benzo[a]pyrene
  • 11. Indeno[1,2,3-c,d]pyrene
  • 12. Dibenzo[a,h]anthracene
  • 13. Benzo[g,h,i]perylene
  • 14. Dibenzo[a,l]perene
  • 15. Dibenzo[a,e]perene
  • 16. Dibenzo[a,i]perene
  • 17. Dibenzo[a,h] perene
Similar Applications