Trastuzumab on bioZen 2.6 µm WidePore C4
LC Conditions
Column
bioZen 2.6 µm WidePore C4, LC Column 100 x 2.1 mm, Ea
Brand
Biozen LC
Part No
00D-4786-AN
Phase Name
WIDEPORE C4
Elution Type
Gradient
Mobile Phase
A: 0.1 % TFA in Water
B: 0.1% TFA in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
70
30
2
5
60
40
3
6
20
80
4
6.1
70
30
5
10
70
30
Flow Rate
0.80 mL/min
Temperature
60°C
System
Waters Waters ACQUITY UPLC I-Class
Detection
LC/UV (DAD, PDA)
Detection Info