EPA Method 537.1 - PFAS in Drinking Water
LC Conditions
Official Method
EPA 537.1
Column
Luna Omega 1.6µm PS C18 100Å, LC Column 100 x 2.1 mm, EA
Brand
Luna Omega
Part No
00D-4752-AN
Phase Name
PS C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Acetic Acid in Water
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
0.5
70
30
3
7
10
90
4
7.5
0
100
5
9
0
100
Flow Rate
0.70 mL/min
Temperature
25°C
System
Agilent Technologies 1290 Infinity Agilent Technologies LC/MS System
Detection
LC/MS/MS
Detection Info
Sample Notes
Drinking Water
Order Items Used in This Application
- 1. PFBS
- 2. PFHxA
- 3. HFPO-DA
- 4. PFHpA
- 5. PFHxS
- 6. ADONA
- 7. PFOA
- 8. PFOS
- 9. PFNA
- 10. 9Cl-PF3ONS
- 11. PFDA
- 12. NMeFOSAA
- 13. PFUnA
- 14. NEtFOSAA
- 15. 11Cl-PF3OUdS
- 16. PFDoA
- 17. PFTrDA
- 18. PFTA
Similar Applications
No Similar applications found.