EPA Method 537.1 - PFAS in Drinking Water

EPA Method 537.1 - PFAS in Drinking Water
LC Conditions
Official Method

EPA 537.1

Column

Luna Omega 1.6µm PS C18 100Å, LC Column 100 x 2.1 mm, EA

Brand

Luna Omega

Part No

00D-4752-AN

Phase Name

PS C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Acetic Acid in Water

B: Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

80

20

2

0.5

70

30

3

7

10

90

4

7.5

0

100

5

9

0

100

Flow Rate

0.70 mL/min

Temperature

25°C

System

Agilent Technologies 1290 Infinity Agilent Technologies LC/MS System

Detection

LC/MS/MS

Detection Info
Sample Notes

Drinking Water

Order Items Used in This Application

Luna Omega 1.6µm PS C18 100Å, LC Column 100 x 2.1 mm, EA
00D-4752-AN

Luna Omega 1.6µm PS C18 100Å, LC Column 100 x 2.1 mm, EA

View Product

  • 1. PFBS
  • 2. PFHxA
  • 3. HFPO-DA
  • 4. PFHpA
  • 5. PFHxS
  • 6. ADONA
  • 7. PFOA
  • 8. PFOS
  • 9. PFNA
  • 10. 9Cl-PF3ONS
  • 11. PFDA
  • 12. NMeFOSAA
  • 13. PFUnA
  • 14. NEtFOSAA
  • 15. 11Cl-PF3OUdS
  • 16. PFDoA
  • 17. PFTrDA
  • 18. PFTA
Similar Applications