Identification and Quantification of Designer Drugs in Urine by LC-MS/MS

Identification and Quantification of Designer Drugs in Urine by LC-MS/MS
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 150 x 3 mm, Ea

Brand

Kinetex

Part No

00F-4462-Y0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: Ammonium formate in water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

6

5

95

3

6.2

5

95

4

8

95

5

Flow Rate

0.50 mL/min

Temperature

50°C

System

Sciex Applied Biosystems Q TRAP,  Mass Analyzer:  QQQ-LT, Interface(s): ES/APCI, PI

Detection

LC/MS/MS

Detection Info
Sample Notes

To 20 µL urine, add 50 µL internal standard (D5-PCP in acetonitrile). Vortex then centrifuge for 10 min at 15,000 rpm. Dilute supernatant with 930 µL Mobile Phase A. Mix and transfer to an autosampler vial for injection.

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 150 x 3 mm, Ea
00F-4462-Y0

Kinetex 2.6 µm C18 100 Å, LC Column 150 x 3 mm, Ea

View Product

  • 1. Dehydronorketamine
  • 2. Norketamine
  • 3. Ketamine
  • 4. Methoxetamine
  • 5. 3-MeO-PCE
  • 6. 3-MeO-PCP
Similar Applications