Identification and Quantification of Designer Drugs in Urine by LC-MS/MS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 150 x 3 mm, Ea
Brand
Kinetex
Part No
00F-4462-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: Ammonium formate in water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
6
5
95
3
6.2
5
95
4
8
95
5
Flow Rate
0.50 mL/min
Temperature
50°C
System
Sciex Applied Biosystems Q TRAP, Mass Analyzer: QQQ-LT, Interface(s): ES/APCI, PI
Detection
LC/MS/MS
Detection Info
Sample Notes
To 20 µL urine, add 50 µL internal standard (D5-PCP in acetonitrile). Vortex then centrifuge for 10 min at 15,000 rpm. Dilute supernatant with 930 µL Mobile Phase A. Mix and transfer to an autosampler vial for injection.
Order Items Used in This Application
- 1. Dehydronorketamine
- 2. Norketamine
- 3. Ketamine
- 4. Methoxetamine
- 5. 3-MeO-PCE
- 6. 3-MeO-PCP
Similar Applications