Fluconazole Test Solution on Kinetex 2.6 µm C18

Fluconazole Test Solution on Kinetex 2.6 µm C18
LC Conditions
Official Method

Ph.Eur. M.2287

Column

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

Brand

Kinetex

Part No

00D-4462-E0

Phase Name

C18

Elution Type

Isocratic

Mobile Phase

A: 0.63g/L Ammonium Formate in water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

17

86

14

Flow Rate

1000.00 mL/min

Temperature

40°C

System

Agilent Technologies Agilent Technologies 1260 Infinity

Detection

LC/UV (DAD, PDA) @ 260 nm (nanometers)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
00D-4462-E0

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

View Product

  • 1. Impurity B
  • 2. Impurity A
  • 3. Fluconazole
Similar Applications

No Similar applications found.