Levothyroxine and Impurity A according to Ph.Eur. Monograph 0401 on Luna Omega 3 µm C18 150 x 4.6 mm

LC Conditions
Official Method

Ph.Eur. M.401

Column

Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Luna Omega

Part No

00F-4784-E0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: Dissolve 1.97 g of phosphoric acid in water and dilute to 2 L with water

B: Dissolve 1.97 g of phosphoric acid in acetonitrile and dilute to 2 L with acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

70

30

2

10

70

30

3

40

20

80

4

50

20

80

Flow Rate

1.00 mL/min

Temperature

25°C

System

Shimadzu Shimadzu Nexera UHPLC

Detection

LC/UV (DAD, PDA) @ 225 nm (nanometers)

Detection Info

Order Items Used in This Application

Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4784-E0

Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Impurity A
  • 2. Levothyroxine
Similar Applications

No Similar applications found.