USP Butamben Impurity Standard Solution 0.2 µg/mL on Kinetex 5 µm XB-C18 100 x 2.1 mm on Agilent 1260 UHPLC
LC Conditions
Column
Kinetex 5 µm XB-C18 100 Å, LC Column 100 x 2.1 mm, Ea
Brand
Kinetex
Part No
00D-4605-AN
Phase Name
XB-C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
1
90
10
3
6
50
50
4
8.5
50
50
5
8.6
90
10
6
10
90
10
Flow Rate
0.40 mL/min
Temperature
30°C
System
Agilent Technologies Agilent Technologies 1200 Series Binary LC System
Detection
LC/UV (DAD, PDA) @ 285 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Aminobenzoic Acid
- 2. Butamben
Similar Applications
No Similar applications found.