Steroid mix on Luna Omega 3 µm C18 150x4.6mm (Batch 1) using Agilent 1260 UHPLC System
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
1.5
80
20
3
16.5
55
45
4
19.5
55
45
5
22.5
80
20
6
25
80
20
Flow Rate
2.00 mL/min
Temperature
30°C
System
Agilent Technologies Agilent Technologies 1260 Infinity
Detection
LC/UV (DAD, PDA) @ 254 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Hydrocortisone
- 2. Dexamethasone
- 3. 11-alpha-hydroxyprogesterone
- 4. 21-hydroxyprogesterone
- 5. 17-alpha-hydroxyprogesterone
- 6. Progesterone
Similar Applications