2023 MSACL poster LLOQ peak 18 OH Corticosterone
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 3 mm, Ea
Brand
Kinetex
Part No
00B-4462-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.5 mM NH4F (aq)
B: MeOH
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
60
40
8
2
50
50
9
4.5
25
75
10
5
5
95
11
6
5
95
12
6.5
60
40
13
8
60
40
Flow Rate
800.00 mL/min
Temperature
30°C
System
Agilent Technologies SCIEX 7500 QTRAP
Detection
LC/MS/MS @ (700°C)
Detection Info
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / well, 96-Well Plates, 2/Pk
Part No
8E-S100-TGB
Condition 1
1 mL MeOH
Sample Pretreatment
Dilute 500 uL serum sample with 500 uL 1% FA.
Load
1 mL total volume of pretreated sample and spike with 10uL IS
Wash 1
1 mL 1% FA
Wash 2
1 mL 30% MeOH
Dry
Dry down under gentle nitrogen flow at 45 degrees C
Elution 1
Twice with 500 uL 1:4 MeOH:ACN
Evaporation
Evaporate to dryness
Reconstitute
100 uL 60:40 0.5mM NH4F (aq) : MeOH
Steps
0
Condition
1 mL MeOH
1
Equilibrate
1 mL Water
2
Load
1 mL total volume of pretreated sample and spike with 10uL IS
3
Wash
1 mL 1% FA
4
Vacuum
Apply vacuum to elute wash
5
Wash
1 mL 30% MeOH
6
Vacuum
Apply vacuum (15-20) for 3-4 minutes until completely dry
7
Elute
Twice with 500 uL 1:4 MeOH:ACN
8
Dry
Dry down under gentle nitrogen flow at 45 degrees C
9
Reconstitute
100 uL 60:40 0.5mM NH4F (aq) : MeOH
Order Items Used in This Application
8E-S100-TGB
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / well, 96-Well Plates, 2/Pk
Similar Applications