USP Teriflunomide Organic Impurities Sensitivity Solution on Kinetex 2.6u C18 150x4.6mm
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Kinetex
Part No
00F-4462-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: Acetonitrile and Buffer (10:90, v/v)
B: Buffer and Acetonitrile (10:90, v/v)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
76
24
2
2.55
76
24
3
12.18
23
77
4
13.14
76
24
5
17
76
24
Flow Rate
1.04 mL/min
Temperature
40°C
System
Waters ACQUITY UPLC H-Class Bio Waters ACQUITY UPLC H-Class
Detection
LC/UV (DAD, PDA) @ 249 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Teriflunomide
Similar Applications