PhEur Atazanavir Sulfate -Impurity K; Reference solution D on Luna 3u C18(2) 150x4.6mm using Agilent 1260 Quaternary UHPLC system
LC Conditions
Column
Luna 3 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna
Part No
00F-4251-E0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: mix 25 volumes of acetonitrile and 75 volumes of a freshly prepared 2.73 g/L solution of potassium dihydrogen phosphate previously adjusted to pH 3.5 with dilute phosphoric acid.
B: mix 25 volumes of a freshly prepared 2.73 g/L solution of potassium dihydrogen phosphate previously adjusted to pH 3.5 with dilute phosphoric acid, and 75 volumes of acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
5.6
95
5
3
8.6
0
100
4
14.6
100
100
5
16
95
5
6
23
95
5
Flow Rate
1.00 mL/min
Temperature
25°C
System
Agilent Technologies 1260 Infinity Agilent Technologies 1200 Series Quaternary LC System
Detection
LC/UV (DAD, PDA) @ 215 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Impurity K
Similar Applications
No Similar applications found.