PhEur Atazanavir Sulfate - Assay; Reference solution C on Luna 3u C18(2) 150x4.6mm using Agilent 1260 Quaternary UHPLC system
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Isocratic
Mobile Phase
A: mix 25 volumes of acetonitrile and 75 volumes of a freshly prepared 2.73 g/L solution of potassium dihydrogen phosphate previously adjusted to pH 3.5 with dilute phosphoric acid.
B: mix 25 volumes of a freshly prepared 2.73 g/L solution of potassium dihydrogen phosphate previously adjusted to pH 3.5 with dilute phosphoric acid, and 75 volumes of acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
50
50
Flow Rate
1.00 mL/min
Temperature
25°C
System
Agilent Technologies 1260 Infinity Agilent Technologies 1200 Series Quaternary LC System
Detection
LC/UV (DAD, PDA) @ 215 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Impurity A
- 2. Impurity C
- 3. Unknown
- 4. Impurity B
- 5. Impurity E
- 6. Impurity F
- 7. Atazanavir
- 8. Impurity G
- 9. Impurity H
Similar Applications
No Similar applications found.