USP Octreotide Acetate Synergy Max-RP - Standard Solution Zoom in

LC Conditions
Column

Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea

Brand

Synergi

Part No

00G-4337-E0

Phase Name

Max-RP

Elution Type

Gradient

Mobile Phase

A: 0.02% (v/v) TFA in Water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

25

65

35

3

30

10

90

4

35

10

90

5

40

90

10

6

45

90

10

Flow Rate

1.00 mL/min

Temperature

40°C

System

Agilent Technologies 1200 Series Agilent Technologies 1200 Series

Detection

LC/UV (DAD, PDA) @ 220 nm (nanometers)

Detection Info
Sample Notes

Standard solution

Order Items Used in This Application

Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
00G-4337-E0

Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Unspecified impurity A
  • 2. Unspecified impurity B
  • 3. Octreotide Acetate
  • 4. Unspecified impurity C
  • 5. Acetyl-Lys5-octreotide
  • 6. Acetyl-Phe1-octreotide
Similar Applications

No Similar applications found.