USP Octreotide Acetate with Synergi Max-RP column - System Suitability Test
LC Conditions
Column
Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Brand
Synergi
Part No
00G-4337-E0
Phase Name
Max-RP
Elution Type
Gradient
Mobile Phase
A: 0.02% (v/v) TFA in Water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
25
65
35
3
30
10
90
4
35
10
90
5
40
90
10
6
45
90
10
Flow Rate
1.00 mL/min
Temperature
40°C
System
Agilent Technologies 1200 Series Agilent Technologies 1200 Series
Detection
LC/UV (DAD, PDA) @ 220 nm (nanometers)
Detection Info
Sample Notes
SST
Order Items Used in This Application
- 1. Octreotide Acetate
- 2. Non-cyclic octreotide
Similar Applications