USP Abacavir and Lamivudine Tablets - Assay (System Suitability Solution)
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: Water and TFA (2000:1)
B: Acetonitrile, Methanol and TFA (1000:1000:1)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
0
2
3.43
100
0
3
10.29
70
30
4
10.37
40
60
5
11.23
40
60
6
11.31
100
0
7
18.31
100
0
Flow Rate
1.50 mL/min
Temperature
40°C
System
Agilent Technologies 1200 Series Agilent Technologies 1200 Series
Detection
LC/UV (DAD, PDA) @ 270 nm (nanometers)
Detection Info
Sample Notes
USP Lamivudine Resolution Mixture C RS (0.8mg) in 2.5mL 0.1N hydrochloric acid
Order Items Used in This Application
- 1. Lamivudine-S-Oxide
- 2. Lamivudine-R-Oxide
- 3. Lamivudine Diasteromer
- 4. Lamivudine
Similar Applications