USP Atazanavir Sulfate (Assay, Organic Impurities – Procedures 1 and 2, Limit of Atazanavir Related Compound A). Atazanavir system suitability solution for USP-Assay
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Isocratic
Mobile Phase
A: Buffer: 2.73 g/L of monobasic potassium phosphate in water. Adjusted pH 3.50 with Orthophosphoric acid (Diluted 1ml to 10 water).
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
15
50
50
Flow Rate
1.00 mL/min
Temperature
25°C
System
Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC H-Class
Detection
UV-Vis Abs.-Diode Array (PDA)
Detection Info
Order Items Used in This Application
- 1. (S,R,S,S)-Diastereomer
- 2. (R,S,S,S)-Diastereomer
- 3. Atazanavir
Similar Applications