USP Amlodipine, Valsartan, and Hydrochlorothiazide Tablets –Assay and Impurities- Amlodipine,Valsartan,Hydrochlorothiazide Standard solution Chromatogram USP-Impurities
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Orthophosphoric acid in water. (1 ml of Orthophosphoric acid in dissolved in 1000 ml of water. Filtered through 0.22µ Nylon filter. Mobile phase A: Acetonitrile: water and Phosphoric acid in ratio of 50:950:1 ml v/v/v.
B: 0.1% Orthophosphoric acid in water. (1 ml of Orthophosphoric acid in dissolved in 1000 ml of water. Filtered through 0.22µ Nylon filter. Mobile phase B: Acetonitrile: water and Phosphoric acid in ratio of 950:50:1 ml v/v/v.
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
3
50
50
3
6
40
60
4
10
5
95
5
10.1
95
5
6
15
95
5
Flow Rate
1.50 mL/min
Temperature
40°C
System
Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC H-Class
Detection
LC/UV (DAD, PDA) @ 210
Detection Info
Order Items Used in This Application
- 1. Benzothiadiazine related Com-A
- 2. Hydrochlorothiazide
- 3. Amlodipine related compound A
- 4. Amlodipine
- 5. Valsartan
Similar Applications