USP Amlodipine, Valsartan, and Hydrochlorothiazide Tablets - Chiral
LC Conditions
Column
Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea
Brand
Lux
Part No
00G-4459-E0
Phase Name
Cellulose-1
Elution Type
Gradient
Mobile Phase
A: n-Hexane, 2-propanol, TFA (850,150,1)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
2
120
100
Flow Rate
0.80 mL/min
Temperature
30°C
System
Agilent Technologies 1260 Infinity Agilent Technologies 1260 Infinity
Detection
LC/UV (DAD, PDA) @ 230 nm (nanometers)
Detection Info
Sample Notes
0.04 mg/mL each of USP Valsartan Related Compound A and USP Valsartan RS in Mobile phase
Order Items Used in This Application
- 1. Valsartan related Compound A
- 2. Valsartan
Similar Applications