USP Amlodipine, Valsartan, and Hydrochlorothiazide Tablets - Chiral

USP Amlodipine, Valsartan, and Hydrochlorothiazide Tablets - Chiral
LC Conditions
Column

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea

Brand

Lux

Part No

00G-4459-E0

Phase Name

Cellulose-1

Elution Type

Gradient

Mobile Phase

A: n-Hexane, 2-propanol, TFA (850,150,1)

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

100

2

120

100

Flow Rate

0.80 mL/min

Temperature

30°C

System

Agilent Technologies 1260 Infinity Agilent Technologies 1260 Infinity

Detection

LC/UV (DAD, PDA) @ 230 nm (nanometers)

Detection Info
Sample Notes

0.04 mg/mL each of USP Valsartan Related Compound A and USP Valsartan RS in Mobile phase​

Order Items Used in This Application

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea
00G-4459-E0

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Valsartan related Compound A
  • 2. Valsartan