USP Amlodipine and Olmesartan Medoxomil Tablets - Standard solution - Kinetex Biphenyl
LC Conditions
Column
Kinetex 5 µm Biphenyl 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Kinetex
Part No
00G-4627-E0
Phase Name
Biphenyl
Elution Type
Gradient
Mobile Phase
A: 6.9 g/L of Sodium phosphate buffer, monobasic, pH 2.5
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
68
32
2
12
68
32
3
15
30
70
4
21
30
70
5
23
68
32
6
25
68
32
Flow Rate
2.00 mL/min
Temperature
60°C
System
Agilent Technologies Agilent Technologies 1200 Series Binary LC System
Detection
LC/UV (DAD, PDA) @ 254 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Amlodipine Besylate
- 2. Olmesartan Medoxomil
Similar Applications