Phenylpropanolmine HCl-Limit of Amphetamine - Modified Method

Phenylpropanolmine HCl-Limit of Amphetamine - Modified Method
LC Conditions
Column

Luna 5 µm C18(2) 100 Å, LC Column 100 x 4.6 mm, Ea

Brand

Luna

Part No

00D-4252-E0

Phase Name

C18(2)

Elution Type

Isocratic

Mobile Phase

A: Methanol

B: THF

C: 2% tetramethylammonium hydroxide and 0.5% Phosphoric acid in water

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

4

0.4

Flow Rate

1.50 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 5 µm C18(2) 100 Å, LC Column 100 x 4.6 mm, Ea
00D-4252-E0

Luna 5 µm C18(2) 100 Å, LC Column 100 x 4.6 mm, Ea

View Product

  • 1. Dextroamphetamine sulfate
  • 2. Phenylpropanolamine
Similar Applications