Environmental Standards
LC Conditions
Column
Ultracarb 5 µm ODS (30) 60 Å, LC Column 150 x 4.6 mm, Ea
Brand
Ultracarb
Part No
00F-0351-E0
Phase Name
ODS (30)
Elution Type
Gradient
Mobile Phase
A: 100% Acetonitrile
B: Acetonitrile/water (20:80)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
0
100
2
4
0
100
3
30.7
80
20
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Furfural
- 2. Aniline
- 3. Phenol
- 4. o-Toluidine
- 5. o-Cresol and m-cresol
- 6. p-Cresol
- 7. N-Methyl aniline
- 8. Nitrotoluene
- 9. Dimethylaniline
Similar Applications