Amino Acids
LC Conditions
Column
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Luna
Part No
00G-4252-E0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: Water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
20
20
80
1
0
100
0
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
Description of Sample Prep and HPLC: Do not use pipette tips or microcentrifuge tubes that have been autoclaved. The water that may collect from the steam and then dry may contaminate the samples. Collection: - Collect sample (3.5-4 ml unheparizied hematocrit tubes of peripheral blood or minimum of 100 mL culture media) - Centrifuge @ 3000 rpm, 4°C, 10 min - Transfer 75mL* of serum or culture media to a new microcentrifuge tube - Sample can now be stored @ -20 °C until ready to perform extraction Extraction: - Add 1.4 mL (2 x 0.7 mL) HPLC-grade methanol, vortex - Spin @ 15,800 g, 0 °C, 15 min - Transfer 1.2mL (2 x 0.6 mL) of supernatant to a new microcentrifuge tube - Dry completely in Speedvac on low heat (2-3 hr) - Store @ -20 °C until ready to run HPLC or proceed directly to next step HPLC: - Warm samples to room temp; resuspend dry extract in 100 mL distilled-deionized water (ddH2O), let sit at RT for about 30 min then vortex - Transfer 95 mL to HPLC sample vials (we use inserts inside vials); no bubbles - HPLC - run on a Luna C18(2) column (Phenomenex, 250mm x 4.6 mm) - 20mL injection volume with a linear water:acetonitrile gradient from 100:0 to 20:80 over 20 min; 1 mL/min - absorbance is read at 225 nm and integrated peaks are calculated by computer - Beckman System Gold 166 detector Standards: 75mL* of each triple standard is extracted as above. Each standard is run in duplicate with the average AUC used to construct a standard curve. Use distilled-deionized water (ddH2O) to make up standards. - Prepare standard stock solutions of kynuerinine, L-tryptophan and 1-methyl-DL-tryptophan (each = 1mg/ml). Protect from light and store @ -70 °C. - Triple standard is a mix of 50ml each of kynuerinine, L-tryptophan and 1-methyl-DL-tryptophan. When brought to a volume of 1ml, the concentration of each = 50 mg/mL - this dilution is used as the most concentrated standard - From the above dilution, make 4 additional serial dilutions (25, 12.5, 6.25 and 1.25 mg/mL) *A 150 mL sample will give a much more readable result if you suspect low amounts of kynuerinine. Directions for dissolving 1-methyl-DL-tryptophan (1 mg/mlL ~4.6mM) total volume 100mL 1-MeTrp 100mg 1N NaOH 1mL distilled H2O 98ml HCl ~1mL Weigh 1-MeTrp and place in bottle. Add 1N NaOH and mix until majority of powder is dissolved. Add water and stir for ~ 1hr. Titrate to pH=7 using 1N HCl. Aliquot and wrap in foil and store @ -70°C.
Order Items Used in This Application
- 1. 5-Hydroxytryptophan
- 2. Kynurenine
- 3. Tryptophan
- 4. N-Methyltryptophan
Similar Applications
No Similar applications found.