Pesticides and Herbicides on Synergi Hydro-RP

Pesticides and Herbicides on Synergi Hydro-RP
LC Conditions
Column

Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea

Brand

Synergi

Part No

00G-4375-E0

Phase Name

Hydro-RP

Elution Type

Gradient

Mobile Phase

A: Water with 0.1% H3PO4

B: Acetonitrile with 0.1% H3PO4

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

30

5

95

Flow Rate

1.50 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, AQ C18 4 x 3.0mm, 10/Pk Part No.: AJ0-7511 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
00G-4375-E0

Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Vacor
  • 2. Prometon
  • 3. Ametryn
  • 4. Prometryn
  • 5. Tebuthiuron
  • 6. Terbutryn
  • 7. Baygon
  • 8. Carbofuran
  • 9. Atrazine / Diuron
  • 10. Bentazon
  • 11. Propazine
  • 12. Barban
Similar Applications

No Similar applications found.