Extraction of Antidepressants on Strata™ X
LC Conditions
Column
Luna 5 µm C8(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna
Part No
00F-4249-E0
Phase Name
C8(2)
Elution Type
Isocratic
Mobile Phase
A: 20 mM KH2PO4, pH 7.0
B: Acetonitrile
C: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
40
30
Flow Rate
2.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Fixed Wave. (UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C8 4 x 3.0mm, 10/Pk Part No.: AJ0-4290 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Pretreatment Note
Blow down to almost dryness under slow stream of N2. Reconstitute with mobile phase to 200 µL. Inject 50 uL on HPLC
Condition 1
1.0 mL Methanol
Condition 2
1.0 mL Water
Load
Load 1.00 mL Sample diluted in Buffer
Wash 1
1.0 mL 5% methanol/D. I. water
Dry
1min at 10'' of Hg
Elution 1
1.0 mL Methanol
Evaporation
Evaporate to dryness
Reconstitute
2 mL Mobile Phase
Order Items Used in This Application
- 1. Nordoxepin
- 2. Nortriptyline
- 3. Doxepin
- 4. Amitriptyline
Similar Applications
No Similar applications found.