Benzodiazepines (SPE-Plasma) on Hybrid C18 Column
LC Conditions
Column
Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea
Brand
Gemini
Part No
00F-4435-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid in Water
B: 0.1% Formic acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
8
20
80
Flow Rate
0.80 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Condition 1
1mL Methanol
Condition 2
1mL Water
Sample Pretreatment
SPE: Strata X 30mg/ml Sample: 0.5ml human plasam + 0.5ml water + 200ng benzodiazepines Condition: 1ml MeOH, 1ml Water Washing: 1ml water, 1ml 40%MeOH in water Elution: 1ml MeOH Reconstitution: 100ul water
Load
Load 1.00 mL Sample diluted in Water
Wash 1
1mL Water
Wash 2
1mL 40% Methanol in Water
Elution 1
1mL Methanol
Evaporation
Evaporate to dryness
Reconstitute
100 µL Water
Order Items Used in This Application
- 1. Chlodiazepoxide
- 2. Flurazepam
- 3. Oxazepam
- 4. Clonazepam
- 5. Temazepam
- 6. Diazepam
Similar Applications