Non-Polar Basics at Low pH on Gemini 5um C6-Phenyl
LC Conditions
Column
Gemini 5 µm C6-Phenyl 110 Å, LC Column 150 x 4.6 mm, Ea
Brand
Gemini
Part No
00F-4444-E0
Phase Name
C6-Phenyl
Elution Type
Isocratic
Mobile Phase
A: 20mM Phosphate buffer pH2.5
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
75
25
Flow Rate
1.50 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C6-Phenyl 4 x 3.0mm, 10/Pk Part No.: AJ0-7915 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Maleic acid
- 2. Chlorpheniramine
- 3. Triprolidine
- 4. Diphenhydramine
Similar Applications