Blood Alcohols on ZB-1701

Blood Alcohols on ZB-1701
GC Conditions
Column

Zebron ZB-1701, GC Cap. Column 30 m x 0.32 mm x 1.00 µm, Ea

Part No

7HM-G006-22

Phase Name

ZB-1701

Injection

Split 13:1 @ 200°C

Carrier Gas

Constant Flow Helium, 7.0 mL/min

Detection

Flame Ionization (FID) @ (220°C)

Detection Info
Sample Notes

Static Headspace: 1mL 50ppm sample in 20mL vial at 70ºC for 15 minutes.

Order Items Used in This Application

Zebron ZB-1701, GC Cap. Column 30 m x 0.32 mm x 1.00 µm, Ea
7HM-G006-22

Zebron ZB-1701, GC Cap. Column 30 m x 0.32 mm x 1.00 µm, Ea

View Product

  • 1. Acetaldehyde
  • 2. Methanol
  • 3. Ethanol
  • 4. Acetone
  • 5. Isopropanol
  • 6. t-Butanol
  • 7. n-Propanol
  • 8. 2-Butanol