EPA 8081 including PCB congeners and o,p'-DDD on ZB-MR1
GC Conditions
Official Method
EPA 8081A
Column
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.32 mm x 0.50 µm, Ea
Part No
7HM-G016-17
Phase Name
ZB-MultiResidue-1
Injection
On-Column @ 103°C
Carrier Gas
Constant Pressure Helium, 3.6 mL/min
Detection
Electron Capture (ECD) @ (340°C)
Detection Info
Sample Notes
Analytes are 50ppb in iso-octane
Order Items Used in This Application
- 1. alpha-BHC
- 2. gama-BHC
- 3. b-BHC
- 4. delta-BHC
- 5. Heptachlor
- 6. BZ# 52
- 7. Aldrin
- 8. Heptachlor epoxide
- 9. gamma-Chlordane
- 10. alpha-Chlordane
- 11. Endosulfan I
- 12. 4,4'-DDE
- 13. 2,4'-DDD
- 14. Dieldrin
- 15. BZ# 118
- 16. Endrin
- 17. 4,4'-DDD
- 18. Endosulfan II
- 19. Endrin aldehyde
- 20. 4,4'-DDT
- 21. Endosulfan sulfate
- 22. Methoxychlor
- 23. Endrin ketone