Gradient Separation of a Cold Mix on Gemini C18

Gradient Separation of a Cold Mix on Gemini C18
LC Conditions
Column

Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea

Brand

Gemini

Part No

00F-4435-E0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 10 mM Phosphate Buffer, pH 2.5

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

15

40

60

3

17

40

60

4

17.01

95

5

5

20

95

5

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Diode Array (PDA) @ (22°C)

Detection Info
Sample Notes

Gradient separation of four analytes in cold remedy on Gemini C18: Phenylephrine, Acetaminophen, Chlorpheniramine and Dextromethorphan and impurity of Maleic acid

Order Items Used in This Application

Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea
00F-4435-E0

Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Phenylephrine
  • 2. Maleic acid
  • 3. Acetaminophen
  • 4. Chlorpheniramine
  • 5. Dextromethorphan
Similar Applications