Pharma ABN Test 12 Mix on Luna 5um C18(2) 150x4.6mm 1mL/min
LC Conditions
Column
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna
Part No
00F-4252-E0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: water w/ 0.1% formic acid
B: acetonitrile w/ 0.1% formic acid
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
1.67
95
5
3
21.69
5
95
4
21.83
95
5
5
30.56
95
5
Flow Rate
1.00 mL/min
Temperature
45°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (25°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Pyridine
- 2. Acetaminophen
- 3. Quinine
- 4. Acebutolol
- 5. Chlorpheniramine
- 6. Triprolidine
- 7. Prednisolone
- 8. 4-Chlorobenzoic acid
- 9. 4-Chlorocinnamic acid
- 10. Diazepam
- 11. Diflunisal
- 12. Hexanophenone
Similar Applications