Pharma ABN Test 12 Mix on Luna 5um C18(2) 150x4.6mm 1mL/min

Pharma ABN Test 12 Mix on Luna 5um C18(2) 150x4.6mm 1mL/min
LC Conditions
Column

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Luna

Part No

00F-4252-E0

Phase Name

C18(2)

Elution Type

Gradient

Mobile Phase

A: water w/ 0.1% formic acid

B: acetonitrile w/ 0.1% formic acid

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

1.67

95

5

3

21.69

5

95

4

21.83

95

5

5

30.56

95

5

Flow Rate

1.00 mL/min

Temperature

45°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (25°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4252-E0

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Pyridine
  • 2. Acetaminophen
  • 3. Quinine
  • 4. Acebutolol
  • 5. Chlorpheniramine
  • 6. Triprolidine
  • 7. Prednisolone
  • 8. 4-Chlorobenzoic acid
  • 9. 4-Chlorocinnamic acid
  • 10. Diazepam
  • 11. Diflunisal
  • 12. Hexanophenone
Similar Applications

No Similar applications found.