Extraction of Carbamate Pesticides using Strata-X on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 5mM Ammonium formate in H2O
B: mM Ammonium formate in ACN
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
97
3
2
0.25
97
3
3
2.5
5
95
4
4
5
95
5
4.01
97
3
6
5.5
97
3
Flow Rate
500.00 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S100-TBJ
Condition 1
1 mL Methanol at 2 mL/min
Condition 2
1 mL 25 mM ammonium acetate pH 5.5 at 2 mL/min
Load
Load 0.40 mL Sample diluted in 1
Wash 1
1 mL 25 mM ammonium acetate pH 5.5 at 2 mL/min
Wash 2
1 mL 15% methanol at 2 mL/min
Dry
2 mins at 10 of Hg
Elution 1
0.5 mL 5% formic acid/methanol at 1 mL/min
Elution 2
0.5 mL 5% formic acid/methanol at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
200 µL 97/3 initial LC mobile phase
Order Items Used in This Application
- 1. Aldicarb-sulfoxide
- 2. Aldicarb-sulfone
- 3. Oxamyl
- 4. Methomyl
- 5. 3-Hydroxy carbofuran
- 6. Aldicarb
- 7. Carbofuran and D3
- 8. Methiocarb
Similar Applications
No Similar applications found.