Pain panel extracted from human urine at cutoff - Kinetex C18 2.6µm 50x2.1mm LCMSMS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid in H2O
B: 0.1% Formic acid in ACN
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
1.3
40
60
3
1.31
5
95
4
1.51
5
95
5
1.52
95
5
6
2.2
95
5
Flow Rate
1.20 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S029-TBJ
Condition 1
1 mL methanol
Condition 2
1 mL 100 mM ammonium acetate buffer (pH 5.95)
Load
Load Sample
Wash 1
2 x 1 mL 100 mM ammonium acetate buffer (pH 5.95)
Wash 2
2 x 1 mL methanol
Dry
5 mins under full vacuum
Elution 1
2 x 0.5 mL 60:24:16 ethyl acetate / IPA / NH4OH
Evaporation
Evaporate to dryness
Reconstitute
1 mL 1 mL Initial Mobile Phase
Order Items Used in This Application
- 1. Morphine
- 2. Morphine-d3
- 3. Oxymorphone
- 4. Hydromorphone
- 5. Hydromorphone-d6
- 6. Codeine
- 7. Codeine-d6
- 8. Naloxone
- 9. Oxycodone
- 10. Oxycodone-d6
- 11. Naltrexone
- 12. Hydrocodone
- 13. cis-Tramadol
- 14. C13-D3-tramadol
- 15. Meperidine
- 16. Meperidine-d4
- 17. Fentanyl
- 18. Buprenorphine
- 19. Buprenorphine-d4
- 20. Methadone
- 21. Methadone-d3
Similar Applications