Intact Antibody on Aeris 3.6um XB-C8 100x4.6mm

Intact Antibody on Aeris 3.6um XB-C8 100x4.6mm
LC Conditions
Column

Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 4.6 mm, Ea

Brand

Aeris

Part No

00D-4481-E0

Phase Name

WIDEPORE XB-C8

Elution Type

Gradient

Mobile Phase

A: 0.1% TFA in Water

B: 0.1% TFA in Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

1

80

20

3

15

40

60

Flow Rate

1.00 mL/min

Temperature

70°C

Detection

UV-Vis Abs.-Diode Array (PDA) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC WIDEPORE C8 for 4.6mm ID, 3/Pk Part No.: AJ0-8771 - Holder Part No.: AJ0-9000

Order Items Used in This Application

Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 4.6 mm, Ea
00D-4481-E0

Aeris 3.6 µm WIDEPORE XB-C8 200 Å, LC Column 100 x 4.6 mm, Ea

View Product

  • 1. Mouse Monoclonal Antibody
Similar Applications